Apertures available for eXact:
There are 4 available apertures for the eXact handheld;
NOTE: These are not changeable in the field.1.5mm, 2.0mm, 4.0mm, 6.0mmThere are 4 available apertures for the eXact handheld;
NOTE: These are not changeable in the field.Aperture & target window size
For each eXact aperture size, the target window size is 1 mm (on radius) larger than the spot measurement area size. This 1 mm over illumination of the measurement area is required by ISO5/4. The white area around the patch, does not influence the measurement data, but does require the operator to center the measurement sample in the target window opening.
Available aperture sizes:
Aperture Size |
Measurement area size |
Opening in target window |
Recommended patch size |
Screening Range |
1.5mm |
1.5mm |
3.5mm |
2mm - 3mm |
175 lines/inch or 69 lines/cm or above |
2mm |
2mm |
4mm |
3mm - 5mm |
133 lines/inch or 52 lines/cm or above |
4mm |
4mm |
6mm |
5mm - 7mm |
65 lines/inch or 26 lines/cm or above |
6mm |
6mm |
8mm |
7mm or larger |
It is always recommended to use the largest aperture size possible.
The illustration below shows how this minimum measurement area appears in relation to the target window opening.
The black circle represents the target window opening.
The white circle represents the actual measurement area (which equals the aperture size).
The magenta square represents the minimum suggested patch size.
This can be used as a guide for operators when placing the instrument optics over patches which do not completely fill the target window opening.