To perform an instrument reset proceed as follows:
Press the two right-hand keys at the same time to ensure you are at the Main CHANNEL MENU.
Next... press the left-most key until you reach the page with the "cal" prompt.
(Note: different 890 series units and some OEM units will have 3-5 pages. In any event, you are seeking the page with the prompt "cal" above the third key).
Here, simultaneously press and release the two center keys (keys 2 & 3).
This puts you into the BIT MENU (Burn In Test).
Press the key beneath the word BURN.
Next press the two center "reset" keys Simultaneously.
Then press the key beneath the word EXIT.
Next press the key beneath the word YES to reset RAM.
After the unit completes the reset it should return to the main menu.
If you are hooked to a printer or software you will need to reprogram the instrument for communication.
**Note is is critical that the densitometer be re-calibrated following a reset,