i7 Reflection Port During Transmission Measurement
Q - Should there be anything in the reflection view port when taking transmission measurements?
A - Yes. The Color i7 transmission kit includes a transmission port plate and a white spectralon disk to be used for transmission calibrations and measurements. The transmission port plate is to be used when doing transmission measurements other than Haze or Reflectance/Transmission. When measuring Haze or Reflectance/Transmission the LAV port plate is used with the transmission spectralon disk at the reflectance port.