EasyTrax IP Address Reset - Communication Failure

EasyTrax IP Address Reset Routine - Communication Failure

1. This routine will restore the Ethernet communication settings back to the factory default:

Addressing Mode: Static

IP Address: 172.16.1.100

Subnet mask: 255.255.0.0

2. To restore factory default settings locate the reset button next to the Ethernet cable jack on the right side of the device. This may be covered by a sticker, which will need to be either moved or removed to gain access.

3. With the instrument powered up, press and hold the reset button for 5 seconds, this will restore the factory communication settings.

NOTE: For more detailed documentation, follow the link to view or download and save this document.

EasyTrax_IP_Address_ Reset.pdf

Haben Sie Fragen? Wünschen Sie ein Angebot? Kontakt(888) 800-9580