Firmware required: Ci6xFirmwareUpdate_r1_1_777.exe
Download the new firmware loader from www.xrite.com/Ci6x/support/d1488
Procedure:
1 | Remove the battery and disconnect both the USB cable and the power cord from the instrument |
2 | Press and hold the measure button while holding the shoe read switch closed (depress the instrument as though you are taking a reading) and connect the power cord. Release both once the display appears white: |
3 | Connect the USB cable |
4 | Start the Frimware Loader - Ci6xFirmwareUpdate_r1_1_777.exe (this is the file downloaded from the link above) |
5 | Click the Get Version button |
6 | The monitor version should appear in the lower left corner |
7 | Click the Upgrade button |
8 | The status of the upgrade will actively update during the process. This process may take several minutes. DO NOT disconnect the USB cable during this process |
9 | "Success" will appear in the lower left corner when the process is complete. |