Power Supply issues

Symptoms of a bad power supply


A defective power supply may cause the following symptoms:

Use only the Switching Power Supply (part number SE30-77) to power the instrument. The SE30-77 requires a line cord. Xrites part number for the appropiate line cord is SD33-07.

 

Power Supplies: The DTP41 Time-Out Error

There have been a number of problems reported related to the DTP41 and power supplies.The correct power supply is X-Rite P/N SE30-77.This is an autoswitching power supply that produces 2.5 amps @ 12 volts. In several instances, customers have tried to use lower-amperage alternatives.The most obvious effect of these low-power situations is an instrument time-out situation.

 

Here is a diagram of the appropriate cables that can be used on the DTP41.

 
 