Model #: 361t

361T

Densitometer

Brak produktu

The 361T Tabletop Transmission Densitometer from X-Rite is ideal for linearizing imagesetters, adjust exposure and performing processor quality control in film processing.



Pobrane



Instrukcje dla użytkownika

Instrukcje dla użytkownika


Dokumentacja techniczna


Specyfikacje

Short Term Repeatability - Density
Ortho,Visual (2 mm and 3 mm aperture):
 0.0 D to 5.0 D: ± 0.01 D
 5.0 D to 5.5 D: ± 2 %
 5.5 D to 6.0 D: ± 3 %
 
Ortho,Visual (1 mm aperture):
 0.0 D to 4.5 D: ± 0.01 D
4.5 D to 5.0 D: ± 1 %
 
UV (3 mm aperture):
 0.0 D to 3.5 D: ± 0.01 D
3.5 D to 4.0 D: ± 1 %
Measurement Geometry Per ANSI PH 2.19, ISO 5/2
Inter-Instrument Agreement 0.0 D - 5.0 D: ± 0.02 D, ± 2 %
Linearity
Ortho,Visual (2mm and 3mm aperture):
0.0 D to 5.0 D: ± 0.02 D
5.0 D to 5.5 D: ± 1%
5.5 D to 6.0 D: ± 2%
 
Ortho,Visual (1mm aperture):
 0.0 D to 4.5 D: ± 0.02 D
4.5 D to 5.0 D: ± 1 %
 
UV (3mm aperture):
0.0 D to 3.5 D: ± 0.02 D
3.5 D to 4.0 D: ± 1 %
Aperture
1mm, 2mm and 3mm
(0.5mm optional)
Humidity ≤ 76%
Operating Temperature Range
10°C to 40°C
50°F to 104°F
Storage Temperature Range
'-20° to 50°C
-4° to 122°F
Dimensions (length, width, height)
43.5 cm x 33.0 cm x 15.2 cm
17.1" x 13.1" x 6.0"
Weight
8.63 kg
19 lbs.
Shipped Weight
17.5 kg
7.95 lbs.
Communication Interface RS-232 serial interface
Instrument Display 8 character LCD
Voltage
90 - 130 VAC 50 - 60 Hz 
120 - 260 VAC 50 - 60 Hz
Warm Up Time
2 minutes
5 minutes (UV)
Wattage 80 VA max.
Density Range
0.0 D to > 6.0 D
0 % to 100 %
Density Standards X-Rite Visual, X-Rite Ortho, X-Rite UV
Industry Standards  ISO 5-2, ANSI PH2.19
Warranty 1 year

Need a Quote? Contact Sales(888) 800-9580