Instrument Settings following a factory reset (SP60, SP62, SP64)
A factory reset will restore the following defaults to the SP60, SP62, SP64 instruments:
- Language - English
-
Measure Options…
- Store Samples - Off
- Pass/Fail - Off
- Auto Std - On
- Averaging - 1
- Diff Display - Off
-
Color Options
- Active Functions… - L*a*b*, L*C*h, and dEcmc
- Active Illum/Obs… - A/10, D65/10, and F2/10
- Opacity Menu
- Data Display - Color
- SPIN k1 - 0.04
- SPIN k2 - 0.60
- SPEX k1 - 0.00
- SPEX k2 - 0.60
- Strength Options
- Method - Apparent
- Predicted - @ 100%
- Metamerism Menu
- Mode - MI (not the DIN method)
- Ill/Obs1 - D65/10
- Ill/Obs2 - F2/10
- dEcmc Factors
- Lightness - 2.00
- Chromaticity - 1.00
- dE94 Factors
- Lightness - 2.00
- Chromaticity - 1.00
- Shade Sort - Off
- SP88 SPEX Mode - Off
-
Database Tools
- View Tags
- In View Tag screen, should be there as all tags were deleted.
-
Hardware Settings
- Serial Port Options
- Baud Rate - 9600
- Handshake - Off
- Auto XMT - Off
- Separator - Comma
- Delimiter - CRLF
- Set Data Types… - Only SPIN Colorimetric should be active
- Header - On
- Std Printout - On
- Emulation - Off
- Read Operation - Switch Only (meaning the switch activated by the foot/shoe)
- Cal Timeout - 24 hrs
- Power Down - 120 sec
- Beeper - Loud
- Clock Adjustments
- Date Format - M/D/Y
- The actual time readings are not affected by factory presets
- Display
- Contrast - 52
- Orientation - Right
- Security - Off
- Error Log… - does not get cleared out, should hopefully be
All of the preceding items can be manually edited via the instrument's "Configuration Menu". The following also occurs:
- All databases will be cleared out, this includes all Samples, Standards, Projects, strings, tags, and jobs.
- L*a*b*, D65/10 and SPIN will all the default selects in QA/Analyze…etc
- Reflectances stored from the last measurement will all be 0's.
- Calibration will be required before any measurements can be done.
- Upon Calibration for the first time after a reset, the black cal will most likely complain about dust, and require a second reading.
Things that will never change due to the Factory Presets being loaded:
- Calibration plaque values in all aperture sizes configured
- Center wave lengths
- Instrument Type (SP60, SP62, SP64), instrument serial #, Cal plaque serial #.
- Initial Cell Strength values (included, excluded, and wall)
- Specular Percents