Reflection Calibration - 8000 Series Benchtop

Reflection Calibration for 8000 Series Benchtop

To calibrate your X-RiteColor Premiere 8000 Series Benchtop instrument for reflection measurements, follow the steps below.  Before beginning the calibration, make sure you have the SRM (Standard Reflectance Material) and the light trap.

SRM

Light Trap

 

To begin calibration, the instrument must be powered up and connected via the USB cable.  For assistance in setting up or connecting your instrument, see Related Support. You must then launch either X-RiteColor Master software or the XR8000 application.  In the  X-RiteColor Master, click on "Instrument" and then "Calibrate".  In the XR8000 application, click on the "Calibrate" button.  Proceed by following the directions on this screen.

 

After the instrument completes the measurement of the SRM, continue by following the directions on the next screen.

 

After the instrument completes the measurement of the light trap, remove the light trap and click on close.  Calibration is then complete.